81
Adjusting the no breath time
If the monitor has the NeoResp license, in the neonatal mode, the 
impedance respiration can set up the detect time for alarm “No 
Breath”.
1. Press the Others key 
2. Select Resp Setup - No Breath Time
The default detection time is 15s.
Adjusting respiration alarms
1. Press the Others key.
2. Select Resp Setup - Resp Rate Alarm.
• To set up the limits, select Adjust Limits.
• To OFF/ON the limits, select RR Alarm.
3. You can also select Alarms Setup to adjust RESP or related 
alarms.
Troubleshooting
•Measurement fails:
− Check the electrode quality and positioning. 
− Adjust the detection limits. During ventilator supported 
breathing, the respiration calculation may count only 
ventilator-produced inspirations and expirations.
− Other electrical devices may interfere with the measurement.