Adjusting the Masimo SpO
2
averaging time
NOTE: For SpO
2
modules with Masimo technology and Masimo
sensors.
1. Press the SpO
2
key.
2. Select Averaging.
Adjusting the Masimo SpO
2
sensor sensitivity level
NOTE: For SpO
2
modules with Masimo technology and Masimo
sensors.
1. Press the
SpO
2
key.
2. Select Sensitivity.
•Use the Normal sensitivity setting for normal patient
monitoring purposes.
•Use the Maximum sensitivity setting for improved poor
perfusion performance and for faster tracking of rapid SpO
2
saturation changes.
Stopping the SpO
2
measurement
1. Remove the SpO
2
sensor from the patient.
2. Disconnect the sensor cable from the monitor.
3. Select Audio Pause key to close the ‘SpO2 probe off’ alarm.
Setting SpO
2
alarms
1. Press the SpO
2
key.
2. Select SpO2 Alarm.
3. Select Adjust Limits to set up the limits.
4. Select SpO2 Alarm to OFF/ON the limits.
5. You can also select Alarms Setup to adjust SpO
2
or related
alarms.