GE MEDICAL SYSTEMS CT9800 QUICK SYSTEM
Rev. 14 Direction 18000
4-3-54
The INITIALIZER PROGRAM Continued:
The initializer program error display is as follow:
PIO ERRORS:
(Same as the Reliability program).
See PIO Errors (later in this section) for more information.
CB ERRORS:
Run time 0.hrs. 0.mins. 2.secs. 0.tenths.
Device Code 24 Unit number 0. State: ***Active***
Modes : - INSTIL UCODE - ZDKINIT
Physical Block : 000000 Sector Count : 42
Memory Address 55206 Command Write
Page Table address 0 Returned XFER count 0
Physical Block: 0
Async Status: 3
CB Execution Error: Hard Errors
CB Status : 100001
Any CB Hard Execution Error
CB Done Bit
CB Error : 40000
DRIVE INTERFACE FAULT
CB Unit Status : 22000
Ready
Port Reserved Bit LSB
Disk Error : 1000
ENTER A COMMAND SELECTION (C=CMD LIST) :
Run Time 0.hrs. 0.mins. 5.secs. 3.tenths.
Device Code 24 Unit number 0. State: ***Active***
Modes : - INSTIL UCODE - ZDKINIT
ERROR WHILE WRITING OUT MICROCODE
To decode this type of error, refer to DG Argus Disk Programmers Reference Series: Models 6236/6237.