ERROR CODE DESCRIPTION
190 Configuration parameter block checksum error
195 Error reading primary magnetic key or checksum error
19A Error reading backup magnetic key or checksum error
19F Magnetic key invalid
1A4 Magnetic key does not match parameter block data & magnetic is active, closed or
error
1A9 Optical key invalid
1AE ODD is turned off. Optical key is returned unit not ready status
1B3 No valid Optical key found. At least one copy returned non blank or checksum error
1B8 Illegal combination of Optical & Magnetic key types; Examine keys
➤ 1BA Magnetic key is set to error state (A7) and optical key is set to active (A4)
➤ 1BD Non-compare of EEPROM facts and magnetic disk facts
1C2 Error reading cache table to Magnetic
1C7 Error writing cache table to Magnetic
1CC Error writing SCRUB directory to magnetic
1D1 Error reading saved DKINIT sectors from magnetic
1D6 Error writing DKINIT sectors to cache
1DB Error backing up cache table following write of DKINIT sectors
1E0 Error writing new keys to optical; less than 2 of 5 successful
1E5 Error writing new platter key to magnetic
1EA Error reading cache table from magnetic fro active platter
1EF Error reading SCRUB directory table from magnetic for active platter
1F4 Error reading SCRUB directory table from completed (closed) platter
INITIALIZATION ERRORS