GE MEDICAL SYSTEMS CT9800 QUICK SYSTEM
Rev. 19 Direction 18000
4-10-21
c. Restore the CT number for Phantoms using the REGEN program at R-level:
R
REGEN ‘CR’
(XENON DETECTOR) (HILIGHT DETECTOR)
- CURRENT CT NUMBERS OF PHANTOMS - - CURRENT CT NUMBERS OF PHANTOMS-
PHANTOM: SMALL MEDIUM LARGE PHANTOM: SMALL MEDIUM LARGE
80KV 0 -94 -94 80KV 0 -134 -134
120KV 0 -94 -94 120KV 0 -100
140KV 0 -94 -94 140KV 0 -94
ENTER KV (80, 120, 140):
NOTE: If this Load-From-Cold is the first post-WC Load-From-Cold, now would be an opportune time to
make a SAVEALL tape for use in future Load-From-Colds or other need for restoration of
customer default data files.
10. Normal Shutdown of operating system (MV7800 and S140):
R
DOWN ‘CR’
!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!
! CT 9800 System shutdown in progress !
! Please stand by . . . !
!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!
(For MV7800) (For S140)
MASTER DEVICE RELEASED MASTER DEVICE RELEASED
AC0AC1 . . . . . . ATU ====> TURN OFF DISK, HIT A KEY <====
0000000000000000 . . ON
(For MV7800, Cont.) (For S140, Cont.)
SCP - CLI> ECLMORTS (MULTI-PROGRAMMING RELIABILITY TEST.
TOTAL # 1K’S - 128. USING MMPU1
SYSTEM TABLES BUILT,
STARTING SUB-SYSTEM SIZING
TEST RUN LIST
TEST # DESCRIPTION