GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM
Rev. 19 Direction 18000
R
REGEN
***REGENERATION OF INSTALLATION INFORMATION***
[RETURN] INDICATES NO CHANGE
CURRENT EXAM NUMBER:
NEW EXAM NUMBER (1-320000):
(XENON DETECTOR) (HILIGHT DETECTOR)
- CURRENT CT NUMBERS OF PHANTOMS - -- CURRENT CT NUMBERS OF PHANTOMS --
PHANTOM: SMALL MEDIUM LARGE PHANTOM: SMALL MEDIUM LARGE
80KV 0 -94 -94 80KV 0 -134 -134
120KV 0 -94 -94 120KV 0 -100 -100
140KV 0 -94 -94 140KV 0 -94 -94
ENTER KV (80, 120, 140):
CURRENT NUMBER OF SCAN DATA FILES:
NEW NUMBER OF SCAN DATA FILES: (can only decrease # of files)
DELETE OVERSCAN OPTION (Y/N): (requires complete Load-from-Cold toreactivate)
CURRENT PHYSICIAN TITLE IS: RADIOLOGIST
PREFERRED PHYSICIAN TITLE, DIAGNOSTICIAN OR RADIOLOGIST (D/R):
CURRENT X-RAY TUBE TYPE
IS:
NOTE: File 6 from the proprietary tape must be loaded prior to selecting D3213T or
D3202T.
ENTER X-RAY TUBE TYPE (1/2/3/4):
(XENON DETECTOR) (HILIGHT DETECTOR)
NEW X-RAY TUBE TYPE: (see WARNING below) NEW X-RAY TUBE TYPE:
1. HIGH PERFORMANCE WITH CUP (D3212T) 1. HIGH PERFORMANCE WITH CUP (D3212T)
2. HIGH PERFORMANCE WITHOUT CUP (D3222T) 2. HIGH PERFORMANCE WITHOUT CUP (D3222T)
3. STANDARD (D1192GR) 3. STANDARD (D1192GR)
4. D3213T 4. D3202T
BE CAREFUL OF TUBE TYPE MISMATCH. DURING YOUR REGEN SECTIONS, THE SYSTEM
SOFTWARE CANNOT DETERMINE WHETHER YOUR SELECTION OF TUBE TYPE IS CORRECT OR
NOT. SELECTION OF THE WRONG TUBE TYPE WILL RESULT IN THE PROGRAM USING THE
WRONG RECONSTRUCTION SOFTWARE. USE OF THE WRONG RECONSTRUCTION SOFTWARE
MAY INTRODUCE IMAGE ARTIFACTS WHICH COULD RESULT IN MISDIAGNOSIS BY THE
PHYSICIAN.
HILIGHT Detector ID is:
ssssssss (8 char max) enter 4 digit detector number, return
New Detector ID:
ssssssss return
HILIGHT detector Assembly Number is:ssssssssssss (12 char max) enter 46 number no dash, return
New Detector Assembly Number:
ssssssssssss return
UNIQUE SYSTEM Number is: ssssssssssssss (14 char max) enter system ID number, return
New System Number:
ssssssssssssss return
WARNING