GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM
Rev. 19 Direction 18000
4-11-4
FASTCAL Selection
IS _____ KV, _____ MM THICKNESS A PREFERRED FASTCAL? (Y/N) (CURRENTLY ____ )
(above prompt repeated 11 times for KV = 80, 120, 140, APERTURE = 1.3, 3.0, 5.0, 10.0 MM)
*** END OF REGENERATION ***
REGEN/S will respond with the following sequence of dialog:
Slope Calibration Logging is: (ON, OFF) (This is current status)
New logging activity value is (Y/N) (Select Y for ON, N for OFF)
Slope Calibration AUTOTEST is: (ON, OFF) (This is current status)
New AUTOTEST activity value is (Y/N) (Select Y for ON, N for OFF)
Exceptions:
If the configuration will not do Slope calibration, an informative message will be typed, and the
subsequent questions will be skipped.
The following message may appear (based on CTSTAT information)
Slope Calibration is only available for HILIGHT detectors
REGEN/A will be normal REGEN, with REGEN/S items inserted before FASTCAL questions.
REGEN/V will output the following data to the user monitor:
Slope Calibration logging will be {ON, OFF} next session
The log limit is nnnnn with {FULL, SHORT, NO QCAL} format
The current log has nnnn entries in {FULL,SHORT,NO QCAL} format
Slope Calibration AUTOTEST will be {ON, OFF} next session
The log limit is nnnnn with {FULL, SHORT, NO QCAL} format
Exceptions:
If the ‘log or autotest’ function if OFF, it’s second line will be skipped
If the log file cannot be OPENED, the third line will be replaced by:
Log file QDR:SLOPE***.LG has been temporarily deleted