GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM
Rev. 12 Direction 18000
6-9A-9
ILLUSTRATION 6-9A-8
1. For a system functional check, perform all of the steps listed in Illustration 6-9A-8.
2. For a previously aligned system, one or more of the steps may be omitted on checkout following replacement
of the tube, collimator or detector. This assumes that the other two components are not altered.
3. Default scans have been setup in Data Collection for most X-Ray alignment scans. Refer to Direction 18008
RDCS Chapter 4, Section 1-2. the default scans have been listed in Illustration 6-9A-8.