GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM
Rev. 12 Direction 18000
6-9A-8
Note: The detector adjustment hardware works such that CW adjustment moves the detector toward the
gantry. This is the opposite of Xenon.
While holding the adjustment stud with a 7/16” wrench, re-tighten the adjustment locking nut. Torque
the inside detector mounting nuts to 25 ft-lbs. When the inside nuts have been torqued, torque the
outside nuts to 25 ft-lbs while holding the inside nut from turning. If a torque wrench is not available
the nuts should be tightened finger tight and then tightened
1/2 flat with a wrench.
9A-3 Procedure Outline
Illustration 6-9A-8 is a protocol flow diagram of the X-Ray alignment and functional check procedures