GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM
Rev. 6 Direction 18000
6-10-5
NOTE: Do not use the black centerline stripe which traverses the length of the cradle
as a reference for Sagittal Light alignment. For most cradles, this stripe will not
pass through the center of the cradle.
(d) Align the Coronal Lights (left and right) along the side white stripe of the tool (Ill. 6-10-2) per section 10.7.
Ideally, the Sagittal Light should intersect both the small hole in the teflon end and the other small hole at
the opposite end near the top adjustment screw.
(e) Align the External Axial Lights (left and right) along the white stripe round the tool at the end where the
mirrors are located (Ill. 6-10-2) per section 10.8.