GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM
Rev. 1 Direction 18000
8-4-2
1X0004 PROBLEM: Inconsistent system data (e.g., illegal device address, partial system
overwirte by user program, etc.).
RECOMMENDATION: Perform a partial initialization.
1X0005 PROBLEM: Master device data error.
RECOMMENDATION: Run disk reliability test.
1X0006 PROBLEM: Master device timeout.
RECOMMENDATION: Run disk reliability test.
1X0007 PROBLEM: Moving head disk error. An address error has occurred and the unit was
not ready. If the unit had been ready, the system would have recalibrated
the unit to resolve the address error.
RECOMMENDATION: Run disk reliability tests.
1X0010 PROBLEM: An undefined interrupt has been detected and it cannot be cleared via an
NIOC. The device code of this device is displayed in the right byte of
AC2.
1X0012 PROBLEM: Insufficient contiguous blocks to build push space indices.
RECOMMENDATION: Perform a full initialization.
1X0013 PROBLEM: Attempted .RTN from level 0 in the background.
RECOMMENDATION: Re-Boot & retry. If problem repeats, run diagnostics & repair hardware.
1X0014 PROBLEM: Inconsistent IPB data (Dual Processors Only).
RECOMMENDATION: Does Not Apply.
1X0015 PROBLEM: Map violation received while control in user interrupt code.
1X0016 PROBLEM: Multi-Bit ERCC error in memory. AC0=DIB, AC1=DIA.
RECOMMENDATION: Refer to “STATE$REP.TM” description in this chapter for further info.
1X0017 PROBLEM: Nova 3/4 Memory Parity Error.
RECOMMENDATION: Does Not Apply.
1X0020 PROBLEM: INFOS Insufficient Memory (INFOS Memory Only).
RECOMMENDATION: Does Not Apply.
1X0021 PROBLEM: Spooler problem (output routine for Line Printer or Console).
RECOMMENDATION: Possible Disk Access Problem. Locate and repair.
1X0022 PROBLEM µ Eclipse Emulator Only.
RECOMMENDATION: Does Not Apply.