1X0023 PROBLEM: Power supply fault (S-280 Only).
RECOMMENDATION: Does Not Apply.
1X0024 PROBLEM: Argus (Acorn) in Dual processor configuration- contention problem.
RECOMMENDATION: Does Not Apply.
1X0070 PROBLEM: PPI has given an invalid interrupt status response or interrupt w/ no status.
RECOMMENDATION: Locate and replace defective PPI.
1X0071 PROBLEM: Interrupt device code “77” detected. This device code does not exist.
System clears each interrupt source from DC=1 until the INTA code 77.
RECOMMENDATION: Possible CPU Backplane jumpering problem. Locate and correct.
1X0077 PROBLEM: During a call to a user mapped space interrupt routine, the software
jumped thru RDOS location 0 (Trap) indicating a CPU or Memory failure.
RECOMMENDATION: Locate and Repair Memory bus failure.
If a system error condition causes a PANIC to occur, bit zero of the panic code word will be reset to zero and the
remainder of the word will contain the system error code. Refer to Appendix A for system error codes.
4-2 - Fortran/RDOS Codes
System errors may also be posted as Fortran (Appendix C) or RDOS codes. They are as follows:
RDOS OR SYSTEM FORTRAN ERROR PARAMETER
ERROR CODE CODE
00000 3 ERFNO Illegal Channel Number
00001 4 ERFNM Illegal File Name
00002 5 ERICM Illegal System Command
00003 6 ERICD Illegal Command for Device
00004 7 ERSV1 Not a save file
00005 8 ERWRO Attempt to Write an Existent File
00006 9 EREOF End of File
00007 10 ERRPR File Read Protected
00010 11 ERWPR File Write Protected
00011 12 ERCRE File Already Exists
00012 13 ERDLE File Does Not Exist
00013 14 ERDE1 Permanent File
00014 15 ERCHA File Attribute Protected
00015 16 ERFOP File Not Open
00016 17 ERFUE Fatal Utility Error
00017 18 EREXQ Execute CL. CM (No Error)
00020 19 ERNUL Invisible Error Code
00021 20 ERUFT Channel Already in Use
00022 21 ERLLI Line Too Long
00023 22 ERRTN Attempt to Restore a Non-Existent Image
00024 23 ERPAR Parity Error
00025 24 ERCM3 Push Depth Exceeded
00026 25 ERMEM Insufficient Memory to Execute Program
00027 26 ERSPC File Space Exhausted
RDOS OR SYSTEM FORTRAN ERROR PARAMETER
ERROR CODE CODE
00030 27 ERFIL File Data Error
00031 28 ERSEL Unit Improperly Selected