GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM
Rev. 3 Direction 18000
8-6-21
*** NO MAIN MEMORY BOARD ***
*** Deadstart Program (DS) *** Revision 1.05
Configuration Information
Processor Type = ST-X
01 Compute Head(s)
02 Back Projector board(s)
ACU Rounding Mode = ST-100
SMP MUX NOT PRESENT
Main Memory 1.0 Mega words
*** RESET Complete ***
PASSED Master ROM test
PASSED Master private RAM test
PASSED Master public memory test
PASSED Master TIMER test
PASSED Slave CP test
PASSED Slave ROM test
PASSED Slave private RAM test
PASSED Slave public memory test
*** RESET Complete ***
Deadstart Part 2 Download in Progress
........
*** FAILED IOP0 test
PASSED PLS Head 0 test
PASSED Backprojector Controller test
failing Back Projectors
00 01 02 03 04 05
Main memory = 0.0 Mega words found
IOP’s are = DAS CHI ---