3-6 S/140 CPU Diagnostics (Continued)
DTOS ONLY
2. Run one pass of diagnostics (typing the underlined):
==> SWREG ’CR’
INPUT IN OCTAL AND ’CR’
SWREG = 000000 001000 ’CR’ (to halt program on error)
==> RUN, 1 ’CR’
DEVICES FOUND BY THE SYSTEM
HOST EQUIPMENT TABLE
EQUIPMENT TABLE #0
NAME CODE
--- 3
TT1 10
TT0 11
RTC 14
LPT 17 (if installed)
MTA 22
ZDKP 27
TT1 50
TT0 51
FPU 76
ACTIVE DIRECTORY = HOST
PROCESSOR UNDER TEST = HOST
LOAD:
PROGRAM NAME: AINST VI REV: 00
PROGRAM DESCRIPTION: NOVA INSTRUCTION SET EXERCISER
PASS 1
LOAD: . . . . (and etc.)
ADES ONLY
2. Run one pass of diagnostics (typing the underlined):
ADES-CLI>RUN (Enter a Carriage Return)
ADES Script Builder REV. 1.0
Building Script
Script Built
Prior X 8.0
NESISV 14.00
NESISV 23.0
NEFADRX 4.0
EASIS_X1 3.0 (Note Underscore. Do not use a space or other character)
EASIS_X2 3.0 (Note Underscore. Do not use a space or other character)
EASIS_X3 3.0 (Note Underscore. Do not use a space or other character)
PEFADRX3.0
E14CHD 2.0
E14IOD 1.0
E14FPUD1 2.0
E14FPUD2 2.0
E14FPUD3 2.0
MMPU1A 2.0
MMPU1B 2.0
E14MMD 2.0
E14ECD 0.0
EMMD_EXER 1.0 (Note Underscore. Do not use a space or other character)
ECLMORT_S 4.0 (Note Underscore. Do not use a space or other character)

Table of Contents