GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM
Rev. 9 Direction 18000
4-3-21
3-6 S/140 CPU Diagnostics (Continued)
DTOS ONLY
The following programs will be run automatically:
AINST V2
EAINST VI
EAINST V2
AINST X (4 passes)
EAINST X1
EAINST X2
EAINST X3
EAINST X4
E14FPD1
E14FPD2
E14FPD3
E14MPD1
E14MPD2
E14MMD
E14ECD
ECLMORTS
At the end of the RUN, 1 sequence, observe:
END RUN 1
If the program has halted on an error in one of the programs, refer to the DTOS Manual Vols. #1 and
#2 for program information and the S/140 manuals for troubleshooting information.
3. Run ECLMORT L overnight (DTOS ONLY- This program is not recognized by ADES. For ADES,
the similar program is called SEARCH which is destructive to disc
contents.)
a. If the DTOS prompt (
==>) is not on the terminal, reboot the DTOS tape as follows:
- Check that the tape is loaded, at BOT (the BOT light is lit), and ONLINE (the READY
light is lit). If tape is loaded, but not at BOT, RESET the ON-LINE switch, press BOT, and
when tape is at BOT, press ON-LINE.
- At the computer, hold the LOCK switch in the UNLOCK position and press RESET.
- At the Virtual Console prompt (!), type:
!100022L
- Observe the diagnostic tape header and type 1 ’CR’ to select the Eclipse, then type 0 ’CR’
to select the S/140.
- Observe the DTOS prompt (
==>)