GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM
Rev. 9 Direction 18000
4-3-22
3-6 S/140 CPU Diagnostics (Continued)
DTOS ONLY
b. Reset the switch register by typing the underlined, as follows:
==> SWREG ’CR’
INPUT IN OCTAL AND ’CR’
SWREG = XXXXXX 000000 ’CR’
==>
If a line printer is available, and a printout is desired, set the switch register to 002000 as follows:
SWREG = XXXXXX 002000 ’CR’
==>
c. Load ECLMORT by typing:
==> LOAD ECLMORT L ’CR’
When the program is loaded an ”@” will be present on the alpha-numeric screen as a prompt.
d. Remove DTOS tape from drive and replace with a scratch tape. When scratch tape is loaded,
ensure tape drive is at BOT, ON-LINE, and READY.
e. Proceed with the program by typing P at the prompt as follows:
@P ’CR’
f. Observe:
ECLMORT L (MULTIPROGRAMMING RELIABILITY TEST, LONG VERSION)
TOTAL # 1K’S = 128 USING MMPUI
SYSTEM TABLES BUILT
STARTING SUB-SYSTEM SIZING DEVICE
SIZING DEVICE 27, DRIVE 0
(Dual Drive operation:
SPACE SELECTS OTHERS DELETE
Sizing Device 27, Drive 0
Sizing Device 27, Drive 1)