GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM
Rev. 9 Direction 18000
4-3-23
3-6 S/140 CPU Diagnostics (Continued)
DTOS ONLY
g. As each test comes up in the following table, the program will wait for you to type a space or
some other character. A space will allow that test to run, while any other character will DELETE
that test. Type a space for each test EXCEPT the line printer test which should be deleted even if
a printer is in the system.
TEST # DESCRIPTION
1 ERROR CORRECTION
2 CHECKER BOARD RANDOM TEST
3 EIS MRI TEST
4 ARITHMETIC TEST
5 ADDRESS TEST
6 EXTENDED ADDRESSING TEST
12 FLT PT TEST
14 LER INSTR/MMPUI ERROR TEST
15 SC MEMORY TEST
20 6061 SERIES DISC (DEV. 27)
21 MAG TAPE TEST (DEV. 22)
23 REAL TIME CLOCK 22 LINE PRINTER (DEV. 17)
24 TTY TEST
ENTER OPTIONS, CR TO CONTINUE
h. Type ‘CR’ to continue.
i. While the program is running, passes and elapsed run time can be observed. To check the run
time, type “4” as follows:
4
R/T (HRS, MIN, ERTOT) = 0 2
This example shows that the program has been running for 2 minutes with no errors. For a pass
and error count, type 7 as follows:
7
R/T = 5
PRG # PASSES
ERRS
1
2
3
4
6
12
14
15
20
23
24