GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM
Rev. 24 Direction 18000
4-3-36
The following is the configuration information for an HTD system with dual ACE drives:0
Enter command (? to see choices): A<CR>
You must have a printer at device code 17 to log your dialogue to a printer.
Do you have such a printer (Y or [N]) ? N<CR>
?
Disk Selection
CT - Computed Tomography C1 - CT using 1 ST34520N
HTD - High Throughput Dynamic
UD - User-defined
ND - No Disk
Select Disk (or EN to END disk selection): HTD<CR>
SCSI DEV.0 (Y or [N]):Y<CR>
Enter BMC priority (0=lowest, 7=highest;S280 and MV4000 0-3 only) [5 ] : <CR>
What is the throttle burst rate [32 ] ? <CR>
What is the throttle break count [4 ] ? <CR>
Current Configuration Facts:
----------------------------
Logical Logical Logical Logical
Unit 0 Unit 1 Unit 2 Unit 3
Heads 1 1 1 1
Sectors 128 128 128 128
Cylinders 8538 611 0 20
Formt’d megs 559 39 0 1
Logical blks 1092614 78001 0 2337
To see the partitioning of a logical unit, enter the unit number: <CR>
MR - Magnetic Resonance
M1 - MR using 1 ST34520N
Microcode on logical unit 0
BMC priority ........... 5
Throttle burst rate .... 32
Break count ............ 4
Enter command (? to see choices): ?U<CR>
Updating...Please wait
SCSI DEV.1(Y or [N]):Y<CR>
Enter BMC priority (0=lowest, 7=highest;S280 and MV4000 0-3 only) [5 ] : <CR>
What is the throttle burst rate [32 ] ? <CR>
What is the throttle break count [4 ] ? <CR>
Current Configuration Facts:
----------------------------
Logical Logical Logical Logical
Unit 0 Unit 1 Unit 2 Unit 3
Heads 1 1 1 1
Sectors 128 128 128 128
Cylinders 8538 611 0 20
Formt’d megs 559 39 0 1
Logical blks 1092614 78001 0 2337
To see the partitioning of a logical unit, enter the unit number: <CR>
Microcode on logical unit 0
BMC priority ........... 5
Throttle burst rate .... 32
Break count ............ 4
Enter command (? to see choices): ?U<CR>
Updating...Please wait
If you have c hanged something and do not wish to have the changes saved use
the [Q]uit command as follows:
Enter command (? to see choices): Q<CR>
EEPROM not updated! Do you really want to abandon these edits (Y/N) ? Y<CR>
After the program has terminated, you may need to do some of the following:
Format or reformat the disk.
Initialize or re-initialize the disk.
Do you want to configure another GESCSI controller (Y/[N]) ?N<CR>
!