GE MEDICAL SYSTEMS CT9800 QUICK SYSTEM
Rev. 24 Direction 18000
4-3-47
Configurator Program Continued:
The CHANGE ALL FACTS command should not be used except to reconfigure to a one or two drive
configuration. The information is already entered into the firmware. If you do reconfigure, you must also write
the changes to EEPROM to save them using the “ U - Update EEPROM” command (this requires switch 2 on the
controller board to be changed to the write enable position “UP” position). Be sure to set this switch back to the
“ DOWN” or write protect position when finished.
THE FOLLOWING IS THE CONFIGURATION INFORMATION FOR A SINGLE ACE DRIVE SYSTEM:
Enter command (? to see choices): A<CR>
You must have a printer at device code 17 to log your dialogue to a printer.
Do you have such a printer (Y or [N]) ? N<CR>
?
Disk Selection
CT - Department CT
MR - Department MR
HTD - High Throughput Dynamic
UD - User-defined
ND - No Disk
Select Disk (or EN to END disk selection): CT<CR>
How many drives in this system (1-4) ? 1<CR> For 1 Disk system
...Disk type(s) accepted.
Enter BMC priority (0=lowest, 7=highest: S280 and MV4000 0-3 only [5 ] : <CR>
...Field accepted.
What is the throttle burst rate [32 ] ? <CR>
What is the throttle break count [4 ] ? <CR>
Current Configuration Facts:
Logical Logical Logical Logical
Unit 0 Unit 1 Unit 2 Unit 3
Heads 1 1 1 1
Sectors 128 128 128 128
Cylinders 4581 0 0 4
Formt’d megs 300 0 0 0
Logical blks 586062 0 0 267
To see the partitioning of a logical unit, enter the unit number:
_________________________________________________________________
Microcode on logical unit 0
BMC priority ............ 5
Throttle burst rate ..... 32
Break count ............. 4
Enter command (? to see choices): ?U<CR>
Updating...Please wait.
...EPROM updated successfully.
If you have changed something and do not wish to have the changes saved use the “ [Q]uit ” command.
Enter command (? to see choices): Q<CR>
EEPROM not updated! Do you really want to abandon these edits (Y/N) ? Y<CR>
After the program has terminated, you may need to do some of the
following:
Format or reformat the disk.
Initialize or re-initialize the disk.
Do you want to configure another GESCSI controller (Y/[N]) ?N<CR>
!
C1 - Department CT using 1 ST34520N
M1 - Department MR using 1 ST34520N