EasyManua.ls Logo

GE D25 - Page 140

GE D25
170 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
D25
User's Manual
GE Grid Solutions
140
994-0081-3.00-21 GE Information
CT/PT Modules, Continued
Procedure:
Over-current
Range
Calibration
(195% - F.S.)
Steps to calibrate the transformer for inputs in the over-current range.
Step Action
1.
Using SGConfig:
Set the calibration and correction factors for the phase and magnitude
to the following default values:
1.00000 for magnitude, and 0.00000 for phase.
To set the calibration factors, go to
D25 AC Configuration>Calibration tab>Internal Calibration tab
To set the correction factors, go to
D25 AC Configuration >Calibration tab>External Correction tab.
Set the configured frequency to match the test signal’s frequency
(either 50 or 60 Hz).
To set the configured frequency, go to:
D25 DCA Configuration>Advanced tab>Line Frequency
Set each input’s magnitude point Report Deadband to zero and
Averaging ON.
To set the Report Deadband, go to:
D25 AC Configuration>I/O Configuration tab>Physical AC
Analog Inputs>Report Deadband
To set the Averaging, go to:
D25 AC Configuration>I/O Configuration tab>Physical AC
Analog Inputs>Averaging
2.
Generate and download the configuration to the unit you are
calibrating.
3.
For each input use a precision AC current source to inject the test
currents indicated in the table in Step 4

Table of Contents

Related product manuals