EasyManua.ls Logo

GE D25 - Page 141

GE D25
170 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE Grid Solutions
D25
User's Manual
GE Information 994-0081-3.00-21
141
CT/PT Modules, Continued
Procedure: Over-current Range Calibration (195% - F.S.) (continued)
Step Action
4.
Using the D25 AC Input Engineering Value Displays (available
through the B050-0 WESMAINT application), record the raw
magnitude values for each input you are calibrating.
Verify that the reported values are within the following tolerances:
CT Type: 1A Part Number: 450-0107
For Test Current… Expected Raw Value is… Tolerance
4.000 A
rms
± 0.05% 8192 ± 200 counts
CT Type: 5A Part Number: 450-0108
For Test Current… Expected Raw Value is… Tolerance
20.000 A
rms
± 0.05% 8192 ± 200 counts
5.
Calculate the 195-FS magnitude calibration factor for each input
using the data gathered in Step 4.
Use the following formula:
Magnitude Calibration Factor = expected value / measured value
For example, if the measured value is 8195, the magnitude
calibration factor is:
8192/8195 = 0.99963
6.
Enter the calibration factors in the Magnitude (195% - FS) for each
input into the configuration (under D25 AC Configuration >
Calibration tab > Internal Calibration tab).
7.
Generate and download the configuration to the unit.

Table of Contents

Related product manuals