K0430 Issue No. 4 12
Pressure Transmitter (P-I) Task
Select the P-I task from the task menu and connect the Unit Under Test (UUT) to the
calibrator as shown below:
A - External pressure source (indicator instruments only) B - Pressure regulator
C - Pressure to current device D - Barrier E - External supply F - Safe Area
• If required, select the output units as described on page 10.
• If applicable, turn on Cal Mode and set-up test parameters as detailed on page 11.
Voltage Output Pressure Transmitter (P-V) Task
Select the P-V task from the task menu and connect the Unit Under Test (UUT) to the
instrument as shown below:
A - External pressure source (indicator instruments only) B - Pressure regulator
C - Pressure to voltage device D - Barrier E - External supply F - Safe Area
• If required, select the output units as described on page 10.
• If applicable, turn on Cal Mode and set-up test parameters as detailed on page 11.
A
B
PRESSURE INT
bar
CURRENT
mA
TASK:P-I
SNAPSHOT MODE
DPI 615 IS
F2F1
EXIT
Max 30V
CAT II
3
P
I
+
-
+
-
C
D
E
D
F
P
V
+
A
B
PRESSURE INT
bar
VOLTAGE
F2F1
EXIT
V
CURRENT
PRESSURE
UNITS
Max 30V
CAT II
TASK: BASIC
DPI 615 IS
CD
3
--
+
+
-
-
C
D
E
F
D
Taking Measurements