K0430 Issue No. 4 42
Test Equipment
The calibration procedures require the following test equipment.
*ppm = parts per million
Table 1 Internal/External Pressure Transducer Verification
Test Equipment and Instrument Parameter/Range
Calibration
Equipment
Accuracy
Calibration
Uncertainty
Digital Voltmeter - 5V input ±30 *ppm ±1 digit ±10 *ppm ±5µV
Digital Voltmeter - 30V input ±45 *ppm ±1 digit ±11 *ppm ± 110 µV
Digital milliammeter - 55mA input ±150 *ppm ±4 digit ±100 *ppm ± 1nA
Digital milliammeter - 24mA output ±150 *ppm ±4 digit ±160 *ppm ± 1nA
Precision thermometer - ambient temperature ±0.2°C ± 0.1°C ±1 digit
Deadweight tester - pressure internal/external Class A deadweight <0.025%
Nominal applied value as a %
of ±full-scale
Permissible deviation
0% (10% for absolute ranges) ± 0.025% FS
20% ± 0.025% FS
40% ± 0.025% FS
60% ± 0.025% FS
80% ± 0.025% FS
100% ± 0.025% FS
80% ± 0.025% FS
60% ± 0.025% FS
40% ± 0.025% FS
20% ± 0.025% FS
0% (10% for absolute ranges) ± 0.025% FS
Calibration

Table of Contents