17 K0430 Issue No. 4
Transmitter Simulator (TX SIM) Task
When used with an external voltage source (see page 24), provides a current output
proportional to the calibrator’s measured output pressure (indicated pressure on
indicator only version). Select task TX SIM. Press EXIT to skip set-up screen if parameters
are correct.
On completion of Tx SIM set-up, the display is configured as follows :.
Connect an external power source to the output loop as detailed on page 24.
To subsequently change any of the Tx SIM scaling parameters, press CHANGE VALUE key
(F1) to obtain the TX Simulation set-up display.
To change the pressure units, press INPUT and select the required scale units. If the
required scale units are not listed, press SETUP, select SETTINGS and proceed as detailed
on page 37.
Taking Measurements