MA-041 DGA 900 Plus - Installation & Commissioning Manual - Rev 1.1 4-Mar-22 Page 60 of 99
Figure 9-36: Configure > Save last measured data > as “short” calibration
Select File > Save As and type the filename calibShort as shown in Figure 9-37.
Figure 9-37: Save As “calibShort”
9.8.2.2.3 Load Test
Connect the 50 Ω resistor across the clips as shown in Figure 9-38.