MA-003 – MULTITRANS User Guide – Rev 3.0  18-Jan-16  Page 23 of 30 
1.  If  the  product  is  in  Standby  mode,  proceed  to  step  3.  If  the  product  is  in 
Measurement mode, manual sampling is prevented. 
2.  Press  the  red  Stop  button  to  stop  the  measurement  process.  The  display  reads 
“OIL  DRAINING”.  If  there  is  oil  in  the  headspace,  the  draining  process  may  take 
several minutes. The product returns to Standby mode. 
3.  When the product is in Standby mode, remove the quick-connect body protector 
by pulling the collar down. 
4.  Fit the male quick-connect fitting and valve assembly to the correct quick-connect 
body on the base of the product as shown in Section 7.1. Ensure that the valve is 
closed to prevent oil leakage. 
5.  Press the white Manual Sample button. The display reads “MANUAL DGA IN: 10s”. 
Subsequent presses of the white button are used to toggle through and select the 
required  oil  circuit  to  sample.  Once  a  circuit  name  is  displayed,  the  timer 
automatically  counts  down  from  10  seconds,  and  if  the  white  button  is  not 
pressed  again,  the  displayed  circuit  is  selected  for  manual  sampling.  Note:  If  the 
product has to first switch oil circuits from the last measured circuit to the selected circuit, an “OIL 
SWITCHING” notice displays. 
6.  An oil line purging process starts. The display reads “MANUAL DGA PURGING OIL 
PRESS KEY WHEN DONE”. 
To  ensure  accurate  results,  oil  must  be  adequately  purged  before  sampling 
begins. This brings fresh oil from the transformer to the manual oil sampling port. 
Typically this means waiting approximately 5–10 minutes, but the recommended 
wait  time  can  vary  depending  on  the  installation  (longer  tubing  and  lower 
temperatures can increase the purge duration). Consult with Technical Support. 
7.  Press  the  white  Manual  Sample  button.  The  display  reads  “{oil  circuit  name} 
MANUAL DGA TAKE OIL SAMPLE PRESS KEY WHEN DONE”. Note: This indicates that the 
oil sampling process can begin. 
8.  Remove the Luer fitting cap, connect the oil sampling equipment e.g. syringe, and 
perform the oil sampling according to your standard process and procedures. 
9.  When  finished  sampling,  close  the  valve  and  disconnect  the  oil  sampling 
equipment. 
10. Remove  the  male  fitting  and  valve  assembly,  and  replace  the  body  protector 
fitting on the quick-connect body. 
11. Press the white Manual Sample  button  to  return the product to  Standby  mode. 
Note:  If  left  in  Manual  Sampling  mode,  the  product  reverts  to  Standby  mode  after  one  hour. 
Measurement then begins at the next scheduled measurement time. 
To start a measurement immediately, press the blue Start button and then select 
the oil circuit to measure. 
To take a manual sample from another oil circuit, press the white Manual Sample button 
again. If a measurement has started since the previous sample, first press the red  Stop 
button. 
See Section 6.1 for example notifications from the LCD panel.