EasyManua.ls Logo

GE LOGIQ C Series - Page 101

GE LOGIQ C Series
164 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE
D
IRECTION 5273008-100, REVISION 9 LOGIQ C SERIES BASIC SERVICE MANUAL
Chapter 4 Functional Checks 4 - 33
4-3-14-5 Full Backup (cont’d)
7.) Press Finish
after Emergency Disk Making complete.
All databases, presets and images should now be saved to Emergency Disk.
8.) Press Ctrl + R, Emergency Disk Recovery window pop-up appear.
9.) Press Next
on pop-up message.
Figure 4-28 Emergency Disk Making Complete