IRECTION 5273008-100, REVISION 9 LOGIQ C SERIES BASIC SERVICE MANUAL
Chapter 8 Replacement Procedures 8-3
Section 8-3Loading Base Image Software
NOTE:While it is believed to be unnecessary, It would not hurt to disconnect the system from the network
and remove all transducers.
NOTE:Please ensure power cable is connected during system upgrade!
1.) Insert the disk labeled “System & Application Software” into the DVD ROM drive.
2.) Properly turn off the scanner by momentarily pressing the Power On/Off Switch. Select “Shut
Down” from the System Exit menu.
3.) If the system will not shutdown normally, hold down the Power On/Off Switch until the light
turns from green to amber.
4.) Turn on the system. System will detect the DVD-RW automatically.
5.) Press any key to continue when below message display as shown below.
Figure 8-2 Upgrade message