IRECTION 5273008-100, REVISION 9 LOGIQ C SERIES BASIC SERVICE MANUAL
10-10 Section 10-4 - Tools Required
10-4-7-2 Dale 600 - Ground Continuity
The Dale 600 measures line cord resistance from the third pin of the attachment plug to the meter’s
Chassis Cable clamp. Test the grounding continuity of the system to all exposed metal parts in
accordance with the IEC 601-1.1 procedure as above. refer to the Dale 600 Instruction Manual for meter
self tests and operation. Record measured resistance of the grounding continuity. The ground wire
resistance should be less than 0.2 (Use any safety analyzer.)