GE HEALTHCARE LOGIQ C3/C5 PREMIUM
DIRECTION 5341787-100, REVISION 8 BASIC SERVICE MANUAL
4 - 26 Section 4-3 - General Procedure
4-3-14 Backup and Restore Database, Preset Configurations and Images
4-3-14-1 Formatting Media
NOTE: The system may crash if use a DVD for Archive/Backup/EZBackup/Restore, so always use a
CD-R or USB HDD for Archive/Backup/EZBackup/Restore.You only can use DVD for upgrade.
1.) To format the backup media, select the UTILITY
button on the Keyboard.
2.) Select CONNECTIVITY
, then REMOVABLE MEDIA. Properly label and Insert the backup
media.
3.) Select the media type from the drop down menu.
4.) Enter the label for the media as shown in Figure 4-34. It is best to use all capital letters with no
spaces or punctuation marks. Press Format
.
5.) The Ultrasound system displays a pop-up menu, as shown in Figure 4-35. When the formatting
has been completed, press OK
to continue.
6.) If desired, verify that the format was successful by returning to Utility
->Connectivity-
>Removable Media
and selecting VERIFY as shown in Figure 4-34.
Figure 4-34 Format and Verify Media

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ C5 Premium and is the answer not in the manual?

GE LOGIQ C5 Premium Specifications

General IconGeneral
BrandGE
ModelLOGIQ C5 Premium
CategoryMedical Equipment
LanguageEnglish

Related product manuals