EasyManua.ls Logo

GE LOGIQ C5 Premium - Scan Probe Environmental Requirements

GE LOGIQ C5 Premium
168 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARE LOGIQ C3/C5 PREMIUM
DIRECTION 5341787-100, REVISION 8 BASIC SERVICE MANUAL
Section 2-2 - General Console Requirements 2 - 5
2-2-4 Scan Probe Environmental Requirements
Operation:10° to 40° C
Storage:-5° to 50° C
NOTE: Temperature in degrees C. Conversion to Degrees F = (Degrees C * (9/5) + 32).
SYSTEMS AND ELECTRONIC PROBES ARE DESIGNED FOR STORAGE TEMPERATURES OF
-10 TO + 60 degrees C. WHEN EXPOSED TO LARGE TEMPERATURE VARIATIONS, THE
PRODUCT SHOULD BE KEPT IN ROOM TEMPERATURE FOR 10 HOURS BEFORE USE.