1 - 12 Section 1-4 - EMC, EMI, and ESD
Section 1-4
EMC, EMI, and ESD
1-4-1 Electromagnetic Compatibility (EMC)
Electromagnetic compatibility describes a level of performance of a device within its electromagnetic
environment. This environment consists of the device itself and its surroundings including other
equipment, power sources and persons with which the device must interface. Inadequate compatibility
results when a susceptible device fails to perform as intended due interference from its environment or
when the device produces unacceptable levels of emission to its environment. This interference is often
referred to as radio–frequency or electromagnetic interference (RFI/EMI) and can be radiated through
space or conducted over interconnecting power of signal cables. In addition to electromagnetic energy,
EMC also includes possible effects from electrical fields, magnetic fields, electrostatic discharge and
disturbances in the electrical power supply.
1-4-2 CE Compliance
The LOGIQ C3/C5 Premium unit conforms to all applicable conducted and radiated emission limits and
to immunity from electrostatic discharge, radiated and conducted RF fields, magnetic fields and power
line transient requirements.
For applicable standards refer to the Safety Chapter in the Basic User Manual.
NOTE: For CE Compliance, it is critical that all covers, screws, shielding, gaskets, mesh, clamps, are in
good condition, installed tightly without skew or stress. Proper installation following all
comments noted in this service manual is required in order to achieve full EMC performance.
1-4-3 Electrostatic Discharge (ESD) Prevention
WARNING
WARNING
DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING
THE NECESSARY ESD PRECAUTIONS:
1.FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC
SENSITIVE EQUIPMENT.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ C5 Premium and is the answer not in the manual?

GE LOGIQ C5 Premium Specifications

General IconGeneral
BrandGE
ModelLOGIQ C5 Premium
CategoryMedical Equipment
LanguageEnglish

Related product manuals