MAINTENANCE
GE Medical Systems
SENOGRAPHE 800T
REV 6 om 2135575–100
3–8
ÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁ
Corrective Actions
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
Explanation / Meaning
of Message
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
Message on Control Console
Message #
ÁÁÁ
ÁÁÁ
I 77
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
COMP PEDAL SECURITY
FAILURE
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
Failure in operator presence
interlock for compression
pedal
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
Switch OFF and switch ON. If fault
continues, note ALL messages and
contact GEMS field service
ÁÁÁ
ÁÁÁ
I 98
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
ROTATION BRAKE OVER-
HEAT
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
The rotation break is hot and
needs to cool down.
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
Wait until the rotation brake tempera-
ture goes down (the message will then
disappear).
ÁÁÁ
ÁÁÁ
I 99
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
AUTO–DECOMPRESSION IS
OFF
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
The automatic decompression
at the end of the exposure has
been disabled in the user select-
able menu.
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
See SET–UP menu programming if
automatic decompression needs to be
activated.
ÁÁÁ
ÁÁÁ
E 01
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
GENERATOR FAILURE
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
Failure in HV section of gene-
rator.
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
Switch OFF and switch ON. If fault
continues, note ALL messages and
contact GEMS field service
ÁÁÁ
ÁÁÁ
E 02
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
FOCUS BIAS FAILURE
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
Failure in tube focus bias cir-
cuits
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
Switch OFF and switch ON. If fault
continues, note ALL messages and
contact GEMS field service
ÁÁÁ
ÁÁÁ
E 03
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
ROTOR FAILURE
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
Failure in anode starter
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
Switch OFF and switch ON. If fault
continues, note ALL messages and
contact GEMS field service
ÁÁÁ
E 05
ÁÁÁÁÁÁÁÁ
FAILURE DURING EXPOSURE
ÁÁÁÁÁÁÁÁ
Current exposure aborted due
to failure in equipment.
ÁÁÁÁÁÁÁÁÁ
Switch OFF and switch ON. If fault
continues, note ALL messages and
contact GEMS field service
ÁÁÁ
ÁÁÁ
E 06
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
SOFTWARE ERROR
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
There is an error in the software
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
Switch OFF and switch ON. If fault
continues, note ALL messages and
contact GEMS field service
ÁÁÁ
ÁÁÁ
ÁÁÁ
E 07
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
SWITCH OFF
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
There is an important prob-
lem in the machine that
requires the user to switch it
OFF.
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
Note ALL messages, switch the
machine OFF, and contact GEMS
field service
ÁÁÁ
ÁÁÁ
E 08
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
POWER SUPPLY FAILURE
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
Supply of generator power cir-
cuits is not authorized
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
Switch OFF and switch ON. If fault
continues, note ALL messages and
contact GEMS field service
ÁÁÁ
E 09
ÁÁÁÁÁÁÁÁ
AEC FAILURE
ÁÁÁÁÁÁÁÁ
The automatic exposure con-
trol is faulty
ÁÁÁÁÁÁÁÁÁ
Switch OFF and switch ON. If fault
continues, note ALL messages and
contact GEMS field service
ÁÁÁ
ÁÁÁ
ÁÁÁ
E 10
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
HEATING FAILURE TRACK 1
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
Large focal spot track is out of
order.
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
Switch OFF and switch ON. If fault
continues, note ALL messages and
contact GEMS field service. It is still
possible to use the small focal spot