MAINTENANCE
GE Medical Systems
SENOGRAPHE 800T
REV 6 om 2135575–100
3–9
ÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁ
Corrective Actions
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
Explanation / Meaning
of Message
ÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁ
Message on Control Console
Message #
ÁÁ
ÁÁ
ÁÁ
E 11
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
HEATING FAILURE TRACK 2
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁ
Small focal spot track is out of
order.
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
Switch OFF and switch ON. If fault
continues, note ALL messages and
contact GEMS field service. It is still
possible to use the large focal spot
ÁÁ
ÁÁ
E 12
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
CHECKSUM ERROR
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁ
Calibration parameters have
been modified.
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
Switch OFF and switch ON. If fault
continues, note ALL messages and
contact GEMS field service
ÁÁ
E 14
ÁÁÁÁÁÁÁÁÁ
CPU or INTERFACE FAILURE
ÁÁÁÁÁÁÁ
The CPU board or the interface
board are out of order.
ÁÁÁÁÁÁÁÁÁÁ
Switch OFF and switch ON. If fault
continues, note ALL messages and
contact GEMS field service
ÁÁ
ÁÁ
E 51
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
COLLIMATOR FAILURE
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁ
The collimator sensors are out
of order.
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
Switch OFF and switch ON. If fault
continues, note ALL messages and
contact GEMS field service
ÁÁ
ÁÁ
E 64
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
FILTER POSITIONING FAIL-
URE
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁ
Filter rotation is out of order, so
good filter positioning is not
ensured. Exposure is inhibited.
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
Switch OFF and switch ON. If fault
continues, note ALL messages and
contact GEMS field service
ÁÁ
ÁÁ
ÁÁ
E 70
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
COMPRESSION FAILURE
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁ
The compression system is out
of order or accurate thickness
calculation is failing in AOP
mode.
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
Switch OFF and switch ON. If fault
continues, note ALL messages and
contact GEMS field service. In case of
thickness failure, the AEC and manual
modes will still be working
ÁÁ
ÁÁ
ÁÁ
ÁÁ
E 71
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
BUCKY FAILURE
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁ
Bucky cannot be moved under
normal conditions.
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
Remove Bucky, then reinstall Bucky.
Make a Rad Prep. Switch OFF and
switch ON. If fault continues, note
ALL messages and contact GEMS
field service. exposures can be made
without the Bucky.
ÁÁ
ÁÁ
E 72
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
COLUMN FAILURE
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁ
Failure in the column system.
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
Switch OFF and switch ON. If fault
continues, note ALL messages and
contact GEMS field service
ÁÁ
ÁÁ
ÁÁ
S 01
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
ABORTED EXPOSURE
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁ
Exposure in progress has been
aborted because exposure but-
ton was released by the opera-
tor before the end of exposure
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
Press RESET and start exam again.
ÁÁ
ÁÁ
ÁÁ
S 02
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
EXPOSURE ABORTED BY HV
ARCING
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁ
Exposure was aborted by
equipment because in excess of
eight X–Ray tube arcings were
produced during the exposure
so far.
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
Press RESET and start exam again.
ÁÁ
ÁÁ
S 03
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
EXPOSURE ABORTED BY
AEC
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁ
Exposure aborted by AEC
because not enough mAs are
available to make exposure.
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
Press RESET, change parameters and
start the exam again.