MAINTENANCE
GE Medical Systems
SENOGRAPHE 800T
REV 6 om 2135575–100
3–10
ÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁ
Corrective Actions
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
Explanation / Meaning
of Message
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁ
Message on Control Console
Message #
ÁÁÁ
S 06
ÁÁÁÁÁÁÁÁ
kV TOO LOW
ÁÁÁÁÁÁÁÁ
kV values are too low to use
AEC (should exceed 24 kV)
ÁÁÁÁÁÁÁÁÁ
Raise kV values.
ÁÁÁ
S 08
ÁÁÁÁÁÁÁÁ
REQUIRED POWER IS TOO
HIGH
ÁÁÁÁÁÁÁÁ
Power requested in manual
mode is too high
ÁÁÁÁÁÁÁÁÁ
Lower kV or mAs values.
ÁÁÁ
S 09
ÁÁÁÁÁÁÁÁ
NO COMPRESSION
DETECTED
ÁÁÁÁÁÁÁÁ
Breast is not compressed
ÁÁÁÁÁÁÁÁÁ
Breast compression is Mandatory in
AOP. Compress the breast.
ÁÁÁ
S 10
ÁÁÁÁÁÁÁÁ
GENERATOR COOLING
xxx s
ÁÁÁÁÁÁÁÁ
Generator requires xxx seconds
to cool
ÁÁÁÁÁÁÁÁÁ
Wait for cooling time to terminate (or
lower mAs values in manual mode).
ÁÁÁ
S 11
ÁÁÁÁÁÁÁÁ
TUBE COOLING xxx s
ÁÁÁÁÁÁÁÁ
Tube requires xxx seconds to
cool.
ÁÁÁÁÁÁÁÁÁ
Wait for cooling time to terminate (or
lower mAs values in manual mode).
ÁÁÁ
ÁÁÁ
ÁÁÁ
ÁÁÁ
ÁÁÁ
ÁÁÁ
S 12
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
OPTICAL DENSITY TARGET
NOT REACHED
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
The target pre–programmed
optical density cannot be
reached. The machine will
make the exposure by allowing
the optical density to go down
to the minimum acceptable
value programmed during
installation.
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
Information message only. no action is
required from the user. The machine
will do the exposure.
ÁÁÁ
S 13
ÁÁÁÁÁÁÁÁ
BACKUP mAs < 50 . EXPO-
SURE INHIBITED
ÁÁÁÁÁÁÁÁ
Maximum mAs <50. exposure
is inhibited.
ÁÁÁÁÁÁÁÁÁ
Wait for cooling time to terminate (or
lower kV values in AEC mode).
ÁÁÁ
ÁÁÁ
ÁÁÁ
S 14
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
MAXIMUM mAs EXCEEDED
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
Maximum mAs set at installa-
tion are exceeded. This mes-
sage appears only if a value
below 600mAs was calibrated
at installation.
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
Lower mAs values.
ÁÁÁ
ÁÁÁ
S 15
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
ARM NOT PRESENT
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
Bad system configuration.
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
Switch OFF, then ON. If the fault con-
tinues, note ALL messages and con-
tact GEMS field service engineer.
ÁÁÁ
ÁÁÁ
S 16
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
SELECTED FILM / SCREEN IS
UNCALIBRATED
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
The selected film / screen pair
is not calibrated.
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
Press the screen / film button to select a
calibrated pair. If the fault continues,
note ALL messages and contact
GEMS field service engineer.
ÁÁÁ
S 25
ÁÁÁÁÁÁÁÁ
TUBE HOUSING OVER TEM-
PERATURE
ÁÁÁÁÁÁÁÁ
The temperature of the tube
housing is too high (> 65°C).
ÁÁÁÁÁÁÁÁÁ
Wait for the tube housing temperature
to fall.
ÁÁÁ
ÁÁÁ
S 26
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
EXPOSURE TOO LONG
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
The exposure is too long. This
message only appears in
manual mode.
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
Increase the kV value and reduce the
mAs value.
ÁÁÁ
ÁÁÁ
ÁÁÁ
S 30
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
COLLIM. / RECEIVER INCOM-
PATIBLE
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
The size of the collimator blade
inserted in the collimator slot,
and the size of the image recep-
tor are incompatible.
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
Check collimator blade size, and insert
correct collimator blade into the colli-
mator slot.