Preparing for a Biopsy
Venue 50 – Basic User Manual 5-27
Direction 5446729-100 English Rev. 7
Displaying the Guidezone (continued)
The needle may vary from the center line or guidezone for
various reasons:
• Needle barrel to needle clearance or strength.
• Bracket manufacturing tolerance.
• Needle deflection due to tissue resistance.
• Needle size chosen. Thinner needles may deflect more.
Table 5-11: Biopsy Guide Availability
Probe Fixed Angle
Multi-Angle
MBX1 MBX2 MBX3
3S-SC 4.0cm 5.5cm 8.0cm
12L-SC 1.5cm 2.5cm 3.5cm
4C-SC 4.1cm 6.07cm 10.05cm
E8CS-SC 15.3 (TR5) 15200
(reusable)
DANGER
Failure to match the guidezone displayed to the guide may
cause the needle to track a path outside the zone.
It is extremely important that when using the adjustable angle
biopsy guides, the angle displayed on the screen matches the
angle set on the guide. Otherwise, the needle will not follow the
displayed guidezone which could result in repeated biopsies or
patient injury.