GE
D
IRECTION GB091046, REVISION 2 VIVID E9 / VIVID E7 BT’13 SERVICE MANUAL
3 - 50 Section 3-8 - Connectivity setup
3-8-8-5 Create new dataflow with Worklist, Database and DICOM Storage
To create a new dataflow with Worklist, Database and DICOM Storage, do the following in exactly these
steps:
1.) Add the Worklist Input, highlight it, select properties, and name it (i.e.: DR PACS Worklist or just
Worklist).
Go to the IP Address field, scroll down and select <Modify>. This will allow you to edit or add IP
Addresses (New Destinations/devices). If the correct IP Address is already in the list, just select it.
2.) Add the Database Input/Output, highlight it, select properties, and name it (i.e.: Local Archive - Int
HD).
Go to the IP Address field and (if the Database is to be the Local Archive) verify that “My Computer”
is selected.
The IP Address should then be correct and the Database Name and Archive File Location entries
should auto-populate correctly.
3.) Add the DICOM Storage Output, highlight it, select properties, and name it (i.e.: DR PACS Store).
Go to the IP Address field, scroll down and select <Modify>. This will allow you to edit or add IP
Addresses (New Destinations/devices).
If the correct IP Address is already in the list, just select it.
If you create the Dataflow in any other order or remove and add an entry later, you may or will receive
“you are not connected to the Database” errors when trying to create a patient locally. The Worklist will
work, however they can NOT save patients “created” on the Vivid 7 to the Archive while using that
dataflow.
You can change the information on each entry in a Dataflow as long as you do NOT remove or add to
the Dataflow out of sequence.
You should also set Direct Search to “All patients” for the new dataflow.
NOTICE
Failure to add destinations to the dataflow in the correct order may or will cause intermittent errors.