IRECTION GB091046, REVISION 2 VIVID E9 / VIVID E7 BT’13 SERVICE MANUAL
Chapter 3 - System setup 3 - 51
3-8-8-6 Create new dataflow with Worklist and DICOM Storage
This dataflow is intended for VIVID E9 / VIVID E7 scanners in a DICOM network, where the data should
not be stored locally on the scanner.
To create a new dataflow with Worklist and DICOM Storage, do the following steps:
1.) Add a new dataflow in Config screen, Connectivity, Dataflow tab. Enter a name, e.g. Worklist -
DICOM Storage.
2.) Select OK.
3.) Add the Worklist input. Configure (at least) IP address, AE title and Port Number.
4.) Select OK.
5.) Add the Dicom Storage output.
• Configure (at least) IP address, AE title and Port Number.
If the site has integrated EchoPAC PC
and CA-1000, enable Allow SR and Allow SR Private Data.
Signed Doppler Velocities is
disabled by default. This means that signs for Doppler Velocities are removed in the DICOM SR.
• See screen dump
6.) Select OK.
7.) If this dataflow should be the default at start-up, enable Default for the dataflow.
Figure 3-37 Dicom SR Settings