GE
D
IRECTION GB091046, REVISION 2 VIVID E9 / VIVID E7 BT’13 SERVICE MANUAL
Chapter 5 - Components and functions (theory) 5 - 17
5-4-1-4 The XY Locking mechanism (cont’d)
The Threaded Lead Screw (4) then rotates CCW and returns to its original position. This is shown in
Figure 5-7 "Unlocked position" on page 5-17.
When the Operator Panel moves into locked position a micro switch (1) is activated.
The micro switch activation is used by the Motor Controller to detect when the Operator Panel is in
locked position.
The XY (frogleg) brakes will stay ON if the VIVID E9 / VIVID E7 is powered down when the Operator
Panel is locked. This enables release of the Park Lock, initiated by the micro switches on the Front
Handle of the Operator Panel.
Figure 5-7 Unlocked position