EasyManua.ls Logo

GE Vivid E7

GE Vivid E7
736 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE
D
IRECTION GB091046, REVISION 2 VIVID E9 / VIVID E7 BT’13 SERVICE MANUAL
5 - 74 Section 5-10 - Power distribution
5-10-4 Power Down Sequence description
5-10-4-1 Overview
There are three possible scenarios for Power Down of the unit:
Power Down
Forced Power Down
Power Loss
Each of the scenarios are described below.
5-10-4-2 Power Down
Press the ON/OFF button (for a short time)
1.) BEP detects the contact of Power (ON/OFF) switch.
2.) PSON_N goes high. This trigs the Main Power Supply to shut down the output voltages.
3.) Controller ACFAIL_N output signal goes low.
4.) Controller TS_OK output signal goes low.
5.) Controller turns OFF the TS (Transmit) voltages.
6.) Controller turns OFF the PMX (Probe MUX) voltages.
7.) Controller turns OFF voltages +/-15, +/-6, +11V, +24V.
8.) ACFAIL_N output signal goes high.
9.) Controller turns OFF +48V.

Table of Contents

Other manuals for GE Vivid E7

Related product manuals