8 - 8 Section 8-4 - Reloading the software
8-4-3 Customer provided prerequisite
• Formatted and labelled media for Images storage.
• Formatted and labelled media for Patient Archive and User Defined Settings.
• Password for the user ADM.
The default password for the user ADM is ulsadm.
If the password has been changed by the site, you should record it for your own use, before you
start the work on the VIVID E9 / VIVID E7.
8-4-4 Tools provided with the Ultrasound scanner at delivery or after an upgrade
• VIVID E9 / VIVID E7 System software (UFD or DVD)
• Application software (UFD or CD)
If present:
• Patch for VIVID E9 / VIVID E7 (UFD or CD)
• Printer Driver Software (UFD or CD)