GE
D
IRECTION GB091046, REVISION 2 VIVID E9 / VIVID E7 BT’13 SERVICE MANUAL
Chapter 8 - Replacement procedures 8 - 13
8-4-9-3 Initiate software reload from the Recovery Console
Ensure that the VIVID E9 / VIVID E7 is powered down.
1.) Depress the On/Off button on the Operator Panel. The VIVID E9 / VIVID E7 starts.
2.) Press the Esc button, on the alphanumeric keyboard, multiple times until the GRUB menu is
displayed on the screen.
3.) On the Grub menu, select: Recovery Console.
The Recovery Console is displayed.
4.) Call the Online Center (OLC).
Tell OLC that you need a “Challenge Code”, and on request, you must read the Challenge String
on the screen. OLC will return a password that you need to type in as the Response.
5.) Enter the Response Code (password) you got from OLC.
Figure 8-2 GRUB menu
Figure 8-3 Recovery Console (Example)
GNU GRUB version 0.97 (638K lower / 2095879K upper memory)
___________________________________________________________
Windows
Recovery Console
Use the | and | keys to select which entry is highlighted.
Press enter to boot the selected OS or 'p' to enter a
password to unlock the next set of features.
USE THE ARROW DOWN KEY
TO SELECT RECOVERY
CONSOLE, THEN PRESS
ENTER.
Enter the password (from OLC).