EasyManua.ls Logo

GE Vivid E7

GE Vivid E7
736 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE
D
IRECTION GB091046, REVISION 2 VIVID E9 / VIVID E7 BT’13 SERVICE MANUAL
8 - 18 Section 8-4 - Reloading the software
8-4-10-2 Installing the Application Software (all BEPs)
Table 8-6 Reloading or loading the Application software from UFD, sheet 1 of 3
Step Task
1.
A menu with “loading…” will appear on the monitor for appr 2
minutes.
A menu with FBReseal will follow.
DO NOT switch off the scanner during this process.
2.
Re-insert the UFD when the “Start Application” dialogue box
appears.
It will take 10-15 seconds until the “Install SW…” button becomes
active.
3.
This dialogue box might not appear.
Select Yes if it appears, the Ultrasound system will then reboot.
4.
The “Install SW….” button has been activated.
Select “Install SW…” to continue.
5. Select OK to continue.

Table of Contents

Other manuals for GE Vivid E7

Related product manuals