8 - 162 Section 8-7 - Replacing XYZ Parts
8-7-1-3 XY Mechanism installation procedure
1.) Position the XY mechanism so it aligns with the holes for the four fixing screws.
2.) Install the four fixing screws.
3.) Install the Lower Frame Assembly.
NOTE: Ensure that Bumper Z enters the slide at the back, before tighten the screws.
4.) Install the Upper Frame Assembly.
5.) Install the Bulkhead Plate.
6.) Install the Bulkhead Board.
7.) Install the Operator Panel, Lower.
8.) Install the Operator Panel, Upper.
9.) Install the LCD Arm assembly.
10.)Install the LCD Monitor assembly.
11.)Install the LCD Rear Cover.
If it is difficult to lock the XY mechanism in parked position, the locking mechanism need adjustments.
Related information:
Section 6-7 "Adjusting the XYZ Mechanism" on page 6-12