GE
D
IRECTION GB091046, REVISION 2 VIVID E9 / VIVID E7 BT’13 SERVICE MANUAL
8 - 246 Section 8-11 - Back End Processor (BEP) parts replacement
8-11-3-5 BEP5 Hard Disk Drive (HDD) installation procedure
Follow these steps to install the Hard Disk Drive:
1.) Slide the HDD bracket hinges into the BEP frame slots.
2.) Connect the cables to the HDD.
3.) Swing the HDD bracket into place.
4.) Install the screw that secures the HDD bracket.
5.) Install the BEP Cover as described in these steps:
a.) Insert the base of the BEP Cover inside the base of the BEP.
Be sure the bottom lip of the BEP Cover rests inside the BEP.
b.) Tilt the top of the BEP Cover toward the VIVID E9 / VIVID E7.
c.) Install the fixing screw(s) for the BEP Cover:
- On BEP6 there are two finger screws on the top of the BEP Cover.
6.) Replace the side cover.
7.) Connect the Mains Power Cable to the wall outlet.
8.) Power up the VIVID E9 / VIVID E7.
9.) Install the System software and the Application software.
10.)Enter the VIVID E9 / VIVID E7’s TCPIP settings and Option strings.
11.)Restore the Patient Archive images, Report Archive, User-Defined Configuration (Customer
Presets), and Service settings.
8-11-3-6 Calibration and adjustments
Calibrate the Front End A/D converters as described in:
• Section 6-6 "DC Offset Calibration (Front End Alignment)" on page 6-11.
Be careful not to pinch any of the cables when installing the BEP Cover.