IRECTION GB091046, REVISION 2 VIVID E9 / VIVID E7 BT’13 SERVICE MANUAL
10 - 12 Section 10-5 - System maintenance
10-5-5 Probe maintenance
10-5-5-1 Probe related checks
10-5-5-2 Basic probe care
The VIVID E9 / VIVID E7 user manuals and various probe handling cards provide a complete
description of probe care, maintenance, cleaning and disinfection. Ensure that you are completely
familiar with the proper care of GE probes.
Ultrasound probes can be easily damaged by improper handling. See the user manual and probe care
cards for more details. Failure to follow these precautions can result in serious injury and equipment
damage. Failure to properly handle or maintain a probe may also void its warranty.
Any evidence of wear indicates the probe cannot be used.
Do a visual check of the probe pins and VIVID E9 / VIVID E7 sockets before plugging in a probe.
TEE and Interoperative probes often have special considerations and individual probe user manuals.
For TEE and Interoperative probes also refer to their separate user manuals.
Table 10-7 Probe Related Checks
Step Item Description
1.
Probe Holders
Clean probe holders. (they may need to be soaked to remove excess gel).
2.
Probes
Thoroughly check the VIVID E9 / VIVID E7 probe connectors and remove dust from inside the
connector sockets if necessary. Visually check for bent, damaged or missing pins.
3. Probes
Verify that the VIVID E9 / VIVID E7 properly recognizes all probes.