IRECTION GB091046, REVISION 2 VIVID E9 / VIVID E7 BT’13 SERVICE MANUAL
Chapter 10 - Care & maintenance 10 - 25
10-6-7-2 Generic procedure on probe leakage current
The most common method of measuring probe leakage is to partly immerse the probe into a saline bath
while the probe is connected to the ultrasound system and active. This method measures the actual
leakage current resulting from the transducer RF drive.
NOTE: Refer to the safety analyzer’s user manual that will be used to perform the tests.