Functional checks
Vivid T9/Vivid T8 – Basic Service Manual 4-33
5795591-100 English Rev. 3
4-3-26-4 ECG Check
Table 4-2: ECG Checks
Step Task Expected Result(s)
1 Connect the ECG harness to the connector
under the control panel.
The unit displays a straight curve along the
bottom edge of the image sector on the screen.
2 Connect the three leads to an ECG simulator,
or
Fasten the three ECG Pads to your body and
connect the three leads to respective ECG
Pad.
When connecting, the signal on the screen will
be noisy.
When the connection is completed, a typical
clean ECG signal is displayed.