Tools required
Vivid T9/Vivid T8 – Basic Service Manual 10-11
5795591-100 English Rev. 3
10-5-3 Special tools, supplies and equipment used for maintenance
Table 10-4: Overview of tool requirements for periodic maintenance
Tool / kit Comments
Digital Volt Meter (DVM)
Anti Static Kit Kit includes anti–static mat, wrist strap and cables for 200
to 240V system
3M #2204 Large adjustable wrist strap
3M #2214 Small adjustable wrist strap
3M #3051 conductive ground cord
Anti Static Vacuum Cleaner 120V
230V
Safety Analyzer The safety Analyzer tool should be calibrated and
compliant with AAMI/ESI 1993 or IEC 60601 or AS/NZS
3551.
QIQ Phantom RMI Grayscale Target Model 403GS
NOTE! The use of a Phantom is not required during
Preventive Maintenance. Customer may use it as part
of their Quality Assurance Program tests.
B/W Printer Cleaning Sheet See printer user manual for requirements
Color Printer Cleaning Sheet See printer user manual for requirements
Disposable Gloves

Table of Contents