System Setup
3-62 Vivid T9/Vivid T8 – Basic Service Manual
5795591-100 English Rev. 3
3-9-6 Query/Retrieve Setup
The Query/Retrieve function makes it possible to search and
retrieve DICOM data from a DICOM server for further analysis
on the Vivid T9/Vivid T8 .
NOTE: You may have to set up Vivid T9/Vivid T8 as a destination on
the server.
3-9-6-1 Query/Retrieve Setup on Vivid T9/Vivid T8
1. Press Config (F2) on the control panel and log on as
administrator.
2. Select Connectivity -> Dataflow.
The Dataflow sheet is displayed.
3. Select Query retrieve from the pull-down menu.
Figure 3-47. Select DICOM Query Retrieve - DICOM Server
Dataflow

Table of Contents